Can continuous sodium ISE measurements stay reliable?

Erik Lindqvist4 min read
Other ManufacturerSensor IntegrationTechnical Reference
Licensed PE Working through this on a live machine? A Maine-licensed engineer can take it from here — included with IMD hardware, by the hour for everything else. Book an engineer

For continuous sodium measurement in a steam-water system, an ion-selective electrode (ISE) can provide the measurement, but electrode drift must be corrected or tracked; a one-time calibration alone does not resolve long-term drift. The decisive quantity is the sodium result after accounting for that drift. The available method identified for doing so is standard addition.

Why one-time calibration and an undefined silica target fail

Calibrating a sodium ISE and then treating its reading as permanently stable leaves electrode drift unaddressed. A continuous indication does not by itself prove that the result remains accurate over time. Use a measurement arrangement that accounts for drift, such as standard addition, and verify the result against the analyzer’s calibration and diagnostic information.

The silica question also needs a defined measurand before selecting an analyzer. The request refers to “oxyde de silice”; the distinction between silica reported as SiO2 and other dissolved silicate forms affects what the measurement represents. Confirm the required reporting basis and the sample conditions with the analytical method or analyzer documentation rather than treating the terms as interchangeable.

Measurement decision What decides it Where to read or confirm
Sodium by continuous ISE Whether electrode drift is being compensated or detected Analyzer method configuration, calibration record, and diagnostic/status information
Silica measurement Whether the required result is SiO2 or another defined silica/silicate basis Plant analytical specification and analyzer method documentation
Hydrazine analysis or substitute Required treatment chemical and analyte measurement method Water-treatment specification and analyzer manufacturer documentation

Sodium ISE response and drift control

An ISE develops an electrical potential that changes with the activity of the target ion. The analyzer converts that potential into a concentration result using its calibration and measurement model. Electrode condition and sample composition affect the measured response, so an electrode can drift even while the analyzer continues to produce a number.

Standard addition addresses this limitation by adding a known amount of the target ion to the sample and observing the response change. The change provides a check on the sample measurement under the actual sample conditions, rather than relying only on a stored calibration. The analyzer’s implementation and calculation details must come from its documentation; do not assume every ISE analyzer performs the same addition sequence or reports drift in the same way.

Checks before selecting a continuous analyzer

  1. Define the required measured species and reporting basis. For silica, state whether the plant needs SiO2 or another form; for sodium, specify that the target is sodium.
  2. Confirm that the analyzer supports continuous service on the intended sample and identify how it compensates for or detects electrode drift.
  3. Review the analyzer’s sample requirements, calibration procedure, standard-addition configuration, and available diagnostic outputs in its manufacturer documentation.
  4. Establish how the continuous result will be checked against the plant’s accepted analytical reference method and how calibration or maintenance events will be recorded.

The mention of Tytronics is a lead to investigate for continuous sodium analyzers, not enough information to select a model or establish its features. Compare the actual analyzer’s documented method and drift-control provisions against the plant requirements.

Hydrazine and silica method selection boundaries

The source request raises hydrazine analyzers and a substitute for hydrazine but provides no analyzer method, product, operating range, or substitute recommendation. Treat these as separate decisions: a hydrazine analyzer measures a defined analyte, while a substitute is a water-treatment chemistry choice. Select neither from the sodium ISE discussion.

For hydrazine, obtain the plant’s treatment specification, required measurement range, sample conditions, and approved analytical method, then match them to analyzer documentation. For silica, first resolve the reporting basis and determine whether the proposed continuous method measures the plant’s intended silica quantity. For both, the method and sample handling must match the process requirement.

Verification after commissioning or service

Verify the sodium channel by checking its calibration status, confirming the standard-addition function or procedure, and comparing the analyzer result with the plant’s reference analysis on a representative sample. Review the trend alongside analyzer diagnostic information: a plausible-looking trend alone cannot distinguish a stable process from a drifting electrode.

For a silica channel, verify that the analyzer method and reported units match the specified SiO2 or other agreed basis, then compare its result to the accepted reference method. For a hydrazine channel, verify that the selected method measures the specified analyte under the actual sample conditions. Record the reference result, analyzer result, calibration state, and any maintenance or reagent changes so later shifts can be separated from instrument drift.

FAQ

Can a sodium ISE measure continuously?

Yes. The method identified for continuous sodium measurement is an ion-selective electrode, but the installation must address electrode drift; standard addition is one method identified for that purpose.

Does standard addition prevent sodium electrode drift?

It provides a way to account for drift by observing the response to a known addition. Check the specific analyzer documentation for how it performs the addition and calculates or flags the result.

Can I select a silica analyzer before defining SiO2?

First confirm whether the required result is silica reported as SiO2 or another defined silica/silicate basis. Match the analyzer method and reporting units to that requirement.

Stop selection or commissioning if the analyzer documentation does not specify its drift handling, sample requirements, or measured/reporting basis. Escalate unresolved method or sample-compatibility questions to the analyzer manufacturer’s official technical support and the plant water-treatment authority.

Back to blog